• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS LUCEA 100; LAMP, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS LUCEA 100; LAMP, SURGICAL Back to Search Results
Model Number ARDLCA219001A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.
 
Event or Problem Description
On 13th december, 2022 getinge became aware of an issue with one of surgical lights - lucea 100.It was stated the bottom support and the bottom cover with handle support were cracked with risk of particles falling.We decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination.
 
Additional Manufacturer Narrative
The correction of h3a device evaluated by manufacturer, h3b device not eval provide code, h3c if other provide code - explain fields deems required.This is based on the internal evaluation.Previous h3a device evaluated by manufacturer: no.Corrected h3a device evaluated by manufacturer: yes.Previous h3b device not eval provide code: other.Corrected h3b device not eval provide code: n/a.Previous h3c if other provide code - explain: device not returned to manufacturer.Corrected h3c if other provide code - explain: n/a.Getinge became aware of an issue with one of surgical lights - lucea 100.It was stated the bottom support and the bottom cover with handle support were cracked with risk of particles falling.We decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination.Based on the information collected, it was established that when the event occurred, the surgical light did not meet its specification and in this way the device contributed to event.Provided information does not indicate if upon the event occurrence, the device was or was not being used for patient treatment.As stated by the subject matter expert at manufacturing site, cracks located on the light head lower covers, at the edge of the on/off button, were detected during daily visual inspection, as recommended by the user manual.If the described failure occurs, the user can visually detect it during the daily checks to be performed prior to each use, or during preventive maintenance.In this case, the user would contact a getinge representative to replace the defective covers of the affected device.For cleaning, the ifu informs the user to not use aggressive and abrasive products.During disinfection, it is prohibited to spray the disinfectant solution directly on the device and to use inappropriate disinfectants.To prevent any incident the lucea 50-100 user manual mentions: ¿check the light heads for chipped paint, impact marks and any other damages¿ during the daily inspections.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
 
Event or Problem Description
Manufacturer's reference number (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUCEA 100
Common Device Name
LAMP, SURGICAL
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key16283003
Report Number9710055-2023-00088
Device Sequence Number20515425
Product Code FTD
UDI-Device Identifier03700712400651
UDI-Public(01)03700712400651
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARDLCA219001A
Device Catalogue NumberARDLCA219001A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2022
Supplement Date Received by Manufacturer06/26/2024
Initial Report FDA Received Date02/02/2023
Supplement Report FDA Received Date06/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/19/2018
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-