Model Number 6900PTFX
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was learned via the implant patient registry that a 6900ptfx 29mm bioprosthetic mitral valve, implanted for six (6) years and two (2) months, was explanted due to torn leaflets.The explanted device was replaced with a 11400m 29mm mitris resilia valve.The patient underwent cabg procedure concomitantly.Per pathology report, the prosthetic valve had three thin focally torn appearing valve cusps.The valve cusps displayed focal yellow fatty streaks.No vegetations identified.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional "customer complaint".The information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
Additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Additional Manufacturer Narrative
Tissue degeneration-related structural deterioration, either calcific or non-calcific, are common chronic failure modes for this type of bioprosthetic heart valve.Operational mechanical stress and biological factors are generally believed to be the major contributors to the non-calcific bioprosthetic tissue degeneration.Structural valve deterioration (svd) can, and typically does, lead to chronic central leaks over a period of time.Svd is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly.Degeneration-related structural deterioration is most commonly related to patient factors and is not usually an indication of a device malfunction.The root cause of this event was determined to be due to patient related factors.The event in this case was impacted by the progression of the patient's underlying valvular disease pathology with structural valve deterioration combined with the patient's other underlying risk factors, including history of coronary artery bypass graft (cabg).The subject device is not available for evaluation as it was discarded.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 - office m2013
irvine, CA 92614
9492506615
MDR Report Key16284062
Report Number2015691-2023-10496
Device Sequence Number4757760
Product Code DYE
UDI-Device Identifier00690103163605
UDI-Public(01)00690103163605(17)200719
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P860057/S022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/19/2020
Device Model Number6900PTFX
Device Catalogue Number6900PTFX29
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/11/2023
Supplement Date Received by Manufacturer03/03/2023
Initial Report FDA Received Date02/02/2023
Supplement Report FDA Received Date03/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/20/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Life Threatening; Hospitalization;
Patient Age67 YR
Patient SexMale
-
-