• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENSEAL G2 STRAIGHT JAW 35CM; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENSEAL G2 STRAIGHT JAW 35CM; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number NSLG2S35
Medical Device Problem Code Difficult or Delayed Activation (2577)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a hemicolectomy at the time of taking tissue in the device does not start the energy cycle.The device is disconnected for a possible failure, it is reconnected and it continues to show the same error.Doctor requests the change of the clamp.The procedure was completed successfully.There were no patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 2/2/2023.Batch #: x95j6x.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was received with no apparent damage.The device was tested with a generator, the hand activation was not functional.However, the device worked properly with the footswitch.The instrument was disassembled to inspect the internal components and it was noted that one of the hand activation wires was disconnected.This caused the hand activation failure.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what caused this issue.A manufacturing record evaluation was performed for the finished device batch x95j6x, and no non-conformances were identified.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 2/24/2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL G2 STRAIGHT JAW 35CM
Common Device Name
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key16286076
Report Number3005075853-2023-00745
Device Sequence Number4881360
Product Code GEI
UDI-Device Identifier10705036014188
UDI-Public10705036014188
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K112033
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNSLG2S35
Device Catalogue NumberNSLG2S35
Device Lot NumberX95P3H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/08/2023
Supplement Date Received by Manufacturer02/07/2023
Initial Report FDA Received Date02/02/2023
Supplement Report FDA Received Date02/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR
Patient SexUnknown
-
-