The complaint is reported as: the nurse checked on a report that a catheter they placed was not functioning.Upon trouble shooting the catheter, it was not flushing, the insertion site was red and irritated.As he withdrew and attempted to flush the saline leaked out of the body of the catheter close to the junction hub.Catheter was removed and patient had other functioning iv.No patient harm was reported.The patient's condition is reported as fine.
|
The complaint is reported as: the nurse checked on a report that a catheter they placed was not functioning.Upon trouble shooting the catheter, it was not flushing, the insertion site was red and irritated.As he withdrew and attempted to flush the saline leaked out of the body of the catheter close to the junction hub.Catheter was removed and patient had other functioning iv.No patient harm was reported.The patient's condition is reported as fine.
|
(b)(4).The customer returned one 6cm x 22ga endurance catheter and syringe for evaluation.Visual inspection of the sample confirmed there was one kink that contained a crack adjacent to the juncture hub.The catheter body also contained a bend.The kink in the catheter body measured approximately 1 mm from the juncture hub.The total length of the catheter body measured 64 mm, (b)(4).The outer diameter of the catheter body measured 0.03525", (b)(4).The catheter was functionally tested per the instructions for use (ifu).The ifu provided with this kit states, "flush catheter using a 10 ml syringe filled with normal saline for injection." when the catheter was flushed with a lab inventory water filled syringe, water leaked from the crack in the kink.The ifu provided with this kit warns the user, "some disinfectants used at catheter insertion site contain solvents which can weaken the catheter material.Alcohol, acetone, and polyethylene glycol can weaken the structure of polyurethane materials.These agents may also weaken the adhesive bond between catheter stabilization device and skin.Allow insertion site to dry completely prior to applying dressing.Do not raise the catheter beyond 90 relative to the skin to avoid catheter kinking and damage." the complaint of an endurance catheter leak was confirmed by a complaint investigation of the returned sample.The catheter body contained one kink with a crack adjacent to the juncture hub.Due to an increased trend in endurance catheter leaks/separations, a non-conformance has been initiated to further investigate this issue.The root cause has not yet been determined.Teleflex will continue to monitor and trend on complaints of this nature.
|