• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW EXT DWELL CATH DEVICE 22GAX6CM; CATHETER INTRAVASCULAR THERAPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW EXT DWELL CATH DEVICE 22GAX6CM; CATHETER INTRAVASCULAR THERAPE Back to Search Results
Catalog Number EDC-00622
Device Problem Fluid/Blood Leak (1250)
Patient Problem Skin Inflammation/ Irritation (4545)
Event Date 12/27/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The complaint is reported as: the nurse checked on a report that a catheter they placed was not functioning.Upon trouble shooting the catheter, it was not flushing, the insertion site was red and irritated.As he withdrew and attempted to flush the saline leaked out of the body of the catheter close to the junction hub.Catheter was removed and patient had other functioning iv.No patient harm was reported.The patient's condition is reported as fine.
 
Event Description
The complaint is reported as: the nurse checked on a report that a catheter they placed was not functioning.Upon trouble shooting the catheter, it was not flushing, the insertion site was red and irritated.As he withdrew and attempted to flush the saline leaked out of the body of the catheter close to the junction hub.Catheter was removed and patient had other functioning iv.No patient harm was reported.The patient's condition is reported as fine.
 
Manufacturer Narrative
(b)(4).The customer returned one 6cm x 22ga endurance catheter and syringe for evaluation.Visual inspection of the sample confirmed there was one kink that contained a crack adjacent to the juncture hub.The catheter body also contained a bend.The kink in the catheter body measured approximately 1 mm from the juncture hub.The total length of the catheter body measured 64 mm, (b)(4).The outer diameter of the catheter body measured 0.03525", (b)(4).The catheter was functionally tested per the instructions for use (ifu).The ifu provided with this kit states, "flush catheter using a 10 ml syringe filled with normal saline for injection." when the catheter was flushed with a lab inventory water filled syringe, water leaked from the crack in the kink.The ifu provided with this kit warns the user, "some disinfectants used at catheter insertion site contain solvents which can weaken the catheter material.Alcohol, acetone, and polyethylene glycol can weaken the structure of polyurethane materials.These agents may also weaken the adhesive bond between catheter stabilization device and skin.Allow insertion site to dry completely prior to applying dressing.Do not raise the catheter beyond 90 relative to the skin to avoid catheter kinking and damage." the complaint of an endurance catheter leak was confirmed by a complaint investigation of the returned sample.The catheter body contained one kink with a crack adjacent to the juncture hub.Due to an increased trend in endurance catheter leaks/separations, a non-conformance has been initiated to further investigate this issue.The root cause has not yet been determined.Teleflex will continue to monitor and trend on complaints of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW EXT DWELL CATH DEVICE 22GAX6CM
Type of Device
CATHETER INTRAVASCULAR THERAPE
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key16291956
MDR Text Key308778721
Report Number9680794-2023-00034
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151513
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEDC-00622
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2023
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/06/2023
Initial Date FDA Received02/03/2023
Supplement Dates Manufacturer Received02/24/2023
Supplement Dates FDA Received02/24/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NOT REPORTED; NOT REPORTED
-
-