Model Number 21-7357-24 |
Device Problem
Failure to Deliver (2338)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/05/2023 |
Event Type
malfunction
|
Event Description
|
It was reported the child have not had any total parenteral nutrition last infusion.Infusion log show that pump has started and the infusion have been on continuous infusion.When pump alarmed reservoir volume low, weight control of infusion bag show that no tpn have infused.No patient injury reported.
|
|
Manufacturer Narrative
|
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.No product was returned; therefore, the reported complaint could not be confirmed and the root cause could not be determined.If the product is returned, this complaint will be reopened for further investigation.
|
|
Manufacturer Narrative
|
Corrected data: h6: health effecs.
|
|