| Model Number |
INT585 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/26/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the implant at tooth location 46 was removed due to fracture at implant body.Additional appointment required: no.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).Patient identifier unknown / not provided.Patient weight unknown / not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).An implant was returned for investigation.Visual inspection of the as returned product identified worn markings due to usage and bone with fracture at the collar.Functional testing to recreate the reported event could not be performed due to the nature of the device and event.No pre-existing conditions were noted on the per.The reported device was located on tooth #46 and was used for approximately 1 year.Pictures or x-ray images were not provided.Review of appropriate documentation: documents reviewed: biomet 3i dental implant ifu (p-iis086gi) rev j - 2022/08 information identified: 'warnings' 'precautions' 'potential adverse events'.Dhr review: dhr review was completed for the subject lot number (2019100865).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review: complaint history review was performed for the reported lot number (2019100865) for similar event and no other complaint was identified.Review completed utilizing keywords: fracture implant post market trend review: feb post market trending was reviewed and there were no actionable events or corrective actions for the reported event or device.Based on the available information, device malfunction did occur and the reported event was confirmed.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|