|
(b)(4).Investigation summary the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pce60a device was returned with no apparent damage and with no reload present.The device was tested for functionality in the straight position with a test reload and achieved its complete firing sequence.The staple line and cut line were complete and the staples meet the staple form release criteria.Upon further inspection, the firing trigger would not function as intended and felt loose.In order to verify the condition of the internal components, the device was disassembled, and the spring firing trigger was noted to be out of its intended position.Based on the information currently available, a product issue was identified during the investigation of the sample received.The damage to the spring firing trigger is potentially related to a manufacturing process.This product issue will be addressed through ethicon endo surgery¿s quality system.A manufacturing record evaluation was performed for the finished device batch number 584a11, and no non-conformances were identified.
|