|
SMITH & NEPHEW, INC. GII POST REF A/P BLK SZ 5; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Model Number |
71440818 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during tka surgery when the surgeon impacted the gii post ref a/p blk sz 5 onto the patient the pins broke off.The procedure was completed, with a non-significant delay, using a s+n back-up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: the device was not returned for evaluation.However, the photograph was reviewed, and revealed that the device is broken.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.With the results of this investigation the root cause of this event could not be determined.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|