An erosion was observed following the implantation of this biotronik device.The sterilization process was investigated.The validated process assures that all sterilization parameters, such as gas concentration, temperature, humidity, etc., are within its specified ranges for each distributed device.Additionally, an analysis of validated microbiological indicators is performed after every sterilization procedure as evidence of successful completion of the sterilization process.Review of the biotronik complaint database did not reveal any changes regarding the trend for this type of incident.In summary, the erosion was not device related.
|
An ulcer was observed following the implantation of this biotronik device.Therefore, the device as received was visually inspected.The visual inspection revealed no external anomalies.The quality documents accompanying the manufacturing process for this device were re-investigated and all production steps were performed accordingly.In particular, the final acceptance test proved the device functions to be as specified.In conclusion there was no indication of a material or manufacturing problem.
|