Model Number 0233050100
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was loss of light.
 
Additional Manufacturer Narrative
Sadditional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Additional Manufacturer Narrative
Alleged failure: cables shorting out or not turning on at all.Salesforce case number (b)(4).The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be bad reed switch, and/or short in the system.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Event or Problem Description
It was reported that there was loss of light.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG., SAFELIGHT FIBER OPTIC CABLE CLEAR, 5MM X 10FT 3.05M
Common Device Name
LIGHT SOURCE, FIBEROPTIC, ROUTINE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key16308858
Report Number0002936485-2023-00085
Device Sequence Number17843299
Product Code FCW
UDI-Device Identifier07613327051650
UDI-Public07613327051650
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K082813
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0233050100
Device Catalogue Number0233050100
Device Lot Number164813
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/11/2023
Supplement Date Received by Manufacturer01/11/2023
Initial Report FDA Received Date02/06/2023
Supplement Report FDA Received Date02/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-