• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P EMPOWR 3D KNEETM INS, 4R 14MM, VE; GUIDE, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P EMPOWR 3D KNEETM INS, 4R 14MM, VE; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Model Number 342-14-704
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 01/23/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to infection.
 
Additional Manufacturer Narrative
The reason for this revision surgery.The agent reported "(the patient has an infection)." the previous surgery and the surgery detailed in this event occurred 11 months and 2 weeks apart.This evaluation is limited in scope as limited information was provided to djo surgical - austin for examination.If information regarding cultures identified in the infection, the severity of the infection or any other relevant information is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR 3D KNEETM INS, 4R 14MM, VE
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, TX 78758
MDR Report Key16309273
Report Number1644408-2023-00122
Device Sequence Number20595138
Product Code FZX
UDI-Device Identifier00888912167512
UDI-Public00888912167512
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K143242
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/12/2024
Device Model Number342-14-704
Device Catalogue Number342-14-704
Device Lot Number138T1039
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2023
Initial Report FDA Received Date02/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
130-03-732 LOT: 155N1990; 241-02-104 LOT: 948R1132; 352-02-104 LOT: 306V1041
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexFemale
-
-