| Model Number |
PM3140 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient presented for a generator change procedure for elective replacement indicator.During the procedure, it was noted that the pacemaker header was failing to connect to the lead.The pacemaker was not used.The physician continued with second pacemaker and completed the procedure.The patient was in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of ¿failure to connect¿ couldn¿t be confirmed.The device was above elective replacement indicator (eri) upon receipt.Device was never implanted.Analysis revealed unsuccessful connection of device is consistent with the type of header that requires the quadripolar lv and two bipolar leads, atrial and right ventricle, that don¿t match the pre-implanted leads.
|
| |
|