| Model Number |
0233050100 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of light.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: cables shorting out or not turning on at all.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be bad reed switch, and/or short in the system.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of light.
|
| |
|