ETHICON INC. SFX SPI PDS+ BI VIO 18IN2 USP1 D/A OS-8; STRATAFIX SPIRAL PDS PLUS BIDIRECTIONAL KNOTLESS TISSUE CONTROL DEVICE Back to Search Results
Catalog Number SXPP2B203
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Trade name - irgacare®, active ingredient(s) ¿ triclosan, dosage form ¿ suture/solid/parenteral, and strength ¿ = 2360 g/m.Component code: g07002 device not returned.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Please provide the lot number.Note: related events reported via 2210968-2023-00877, 2210968-2023-00878, and 2210968-2023-00880.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and a barbed suture was used.During the procedure, the needle popped off the suture while suturing capsule closed.It was unknown if he procedure was completed successfully.There were no adverse patient consequences.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 3/6/2023.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional test evaluation were conducted on the returned devices.Visual analysis of the returned sample determined that twenty-two packets of product code sxpp2b203 were received for evaluation.Upon initial inspection, of the samples, no external damages were observed on the packets.As per the sampling plan, a visual inspection was performed on thirteen samples, the packets were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand no anomalies were observed during the evaluation.A functional test was performed using instron equipment and the pull force result was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SFX SPI PDS+ BI VIO 18IN2 USP1 D/A OS-8
Common Device Name
STRATAFIX SPIRAL PDS PLUS BIDIRECTIONAL KNOTLESS TISSUE CONTROL DEVICE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km. 8.3
san lorenzo 00754
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key16310866
Report Number2210968-2023-00879
Device Sequence Number17879261
Product Code NEW
UDI-Device Identifier10705031465169
UDI-Public10705031465169
Combination Product (Y/N)Y
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K192144
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Catalogue NumberSXPP2B203
Device Lot NumberSMBAEL
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2023
Supplement Date Received by Manufacturer02/16/2023
Initial Report FDA Received Date02/07/2023
Supplement Report FDA Received Date03/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-