| Model Number |
N/A |
| Medical Device Problem Codes |
Unstable (1667); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Osteolysis (2377)
|
| Date of Event |
12/28/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported the patient underwent a right knee revision approximately thirteen months post-implantation due to pain and loosening of the tibial tray.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical devices: femur cemented cruciate retaining (cr) standard right size 11 catalog#: 42502607002 lot#: 64355854.Articular surface medial congruent (mc) right 10 mm height catalog#: 42522100910 lot#: 64798941.Unknown bone cement catalog#: ni lot#: ni.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as its location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2023-00014, 3007963827-2023-00031, and 0001822565 -2023-00464.D10 medical devices: femur cemented cruciate retaining (cr) standard right size 11, catalog#: 42502607002, lot#: 64355854.Articular surface medial congruent (mc) right 10 mm height, catalog#: 42522100910, lot#: 64798941.Unknown bone cement catalog#: ni, lot#: ni.Unknown tibial stem extension catalog#: ni, lot#: ni.
|
| |
|
Event or Problem Description
|
|
It was reported the patient underwent a right knee revision approximately thirteen months post-implantation due to pain and loosening of the tibial tray.During the revision, bone loss was noted around the femoral and the tibial tray and stem.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Updated: b4, g3, g6, h2, h3, h6, and h10.Reported event was confirmed by review of medical records.Medical records and radiographs were provided and reviewed by a health care professional.Review of the available medical records identified the following: radiolucent lines along the tibial component, aseptic loosening of the tibial component, tibial and femoral bone loss and instability, and excessive scar tissue surrounding patella.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|