COOK INC ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER; GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number N/A
Medical Device Problem Codes Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.Pma/510(k) #: exempt this report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported the physician was experiencing difficulty using the ultrathane mac-loc locking loop multipurpose drainage catheter.The physician demonstrated to a cook representative how he was experiencing difficulty advancing and removing the flexible stiffener from the catheter.The physician pulled the device off the shelf and showed the device "getting stuck" when both unflushed and flushed.No patient was present during the demonstration.No adverse effects were reported for this incident,.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unchanged, unknown, or unavailable.Investigation ¿ evaluation on 02feb2023, cook medical received a complaint from a representative of christus schumpert highland, shreveport, la (united states).They reported the flexible stiffener from an ultrathane mac-loc locking loop multipurpose drainage catheter (rpn: ult12.0-38-45-p-6s-clm-rh; lot: 15064002) was difficult to advance and remove from the catheter.No patient contact was made with device, as this device was used by a physician to demonstrate the issue he was encountering.He showed both unflushed and flushed where stiffener was getting stuck.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions, quality control procedures, and specifications, as well as a visual inspection, functional test, and dimensional verification pf the returned complaint device, were conducted during the investigation.The device was returned in an opened and prior to use condition.Upon wiping off the supplied flexible stiffener with a moist towelette, the stiffener was able to be inserted and removed from the drainage catheter without encountering difficulty.Visual examination of both devices confirmed there was no damage.A dissection of the catheter was performed, confirming the inner diameter was within specification.Additionally, the outer diameter of the flexible stiffener was also confirmed to be within specification.Additionally, a document-based investigation evaluation was performed.In response to this incident, cook completed a review of the product device master record (dmr) and concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Cook also reviewed the device history record (dhr).The dhr for lot 15064002 and related subassembly lots confirmed one relevant recorded non-conformance for "curve not to template".These devices were scrapped prior to further processing of the order.To date, a further search of our database records revealed one additional complaint associated to the reported lot from the same facility.Cook also reviewed product labeling.The product ifu, [t_multi2] ¿multipurpose drainage catheter,¿ provides the following information to the user related to the reported failure mode: precautions: ¿when inserting a stiffening cannula into a catheter with retention suture, hold suture during cannula insertion to avoid bunching or tangling of suture.¿ how supplied: ¿supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened or undamaged.Do not upset he product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.¿ the information provided upon review of the dmr, ifu, dhr and returned device suggests that the device was not manufactured out of specification, and that there are no nonconforming devices in house or out in the field.Based on the information provided, inspection of the returned device, and the results of the investigation, cook medical has concluded that a component failure without a manufacturing or design defect contributed to this incident.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER
Common Device Name
GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key16312379
Report Number1820334-2023-00102
Device Sequence Number7049269
Product Code GBO
UDI-Device Identifier00827002097674
UDI-Public(01)00827002097674(17)251110(10)15064002
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/10/2025
Device Model NumberN/A
Device Catalogue NumberULT12.0-38-45-P-6S-CLM-RH
Device Lot Number15064002
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/02/2023
Supplement Date Received by Manufacturer08/03/2023
Initial Report FDA Received Date02/07/2023
Supplement Report FDA Received Date08/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-