Model Number 1456Q/86
Device Problem Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/30/2023
Event Type  malfunction  
Event Description
It was reported during an implant procedure on (b)(6) 2023, the left ventricular lead stylet broke inside the cable.The lead was replaced within the same operation.Post-procedure there were no patient consequences.
 
Manufacturer Narrative
Correction: return not received, appropriate codes inputted.
 
Manufacturer Narrative
The reported event of ¿the stylet broke the inside cable¿ was confirmed.As received, a complete lead with stylet stuck was returned in one piece.The ptfe coating of the stuck stylet was stripped and was found bunched up with the inner coil distal to the connector pin.The connector pin and crimp sleeve were found pulled out of the connector assembly stretching the inner coil consistent with damage due to excessive forces applied while attempting to remove the stylet from the lead during the procedure.The connector cap was in place and intact in the connector assembly.The cause of the reported event was isolated to the bunching of the stylet ptfe coating inside the inner coil at the connector region that prevented the removal of the stylet and excessive forces resulted in the connector pin to be pulled out of the connector assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD SMALL-S, 86 CM
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16313601
MDR Text Key308962317
Report Number2017865-2023-05304
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510158
UDI-Public05414734510158
Combination Product (y/n)Y
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 07/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1456Q/86
Device Catalogue Number1456Q/86
Device Lot NumberA000128928
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/06/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-