| Model Number |
9505-02-801 |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial , a follow-up will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported on the box for tray 2, several plastic parts came of the box when it was washed and the notch impactor is way to scratched.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : the device associated with this report was not returned to depuy synthes for evaluation.The photo investigation confirmed the reported allegation.The photographs provided show broken material from the prep case.However foreign substance or debris on the case cannot be seen.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the investigation was not able to retrieve the required lot code.
|
| |
|
Search Alerts/Recalls
|