| Model Number |
C012D |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Investigation summary: the product was returned for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that one empty open overwrap packet, one paper lid and one winding former with seven needle-suture combination, a suture and one detached needle that pertain to product code c012d.This product code is multistrand and required eight needle-suture per packet.After an investigation of the suture, no missing suture was observed.However, short insertion was noted in the strand.The visual inspection concluded that the combination of the needle/suture was detached from the needle since the insertion end of the suture did not meet the requirement.Based on the information currently available, short insertion issue was identified as pull out during the investigation of the sample received.This product issue will be addressed through the ethicon quality system.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a an unknown procedure on an unknown date and suture was used.Before use on the patient, it was found that one suture had been missing.Visual analysis of the returned sample determined that one empty open overwrap packet, one paper lid and one winding former with seven needle-suture combination, a suture and one detached needle that pertain to product code c012d.This product code is multistrand and required eight needle-suture per packet.After an investigation of the suture, no missing suture was observed.However, short insertion was noted in the strand.The visual inspection concluded that the combination of the needle/suture was detached from the needle since the insertion end of the suture did not meet the requirement.No adverse patient consequences were reported.
|
| |
|
Search Alerts/Recalls
|
|
|