| Model Number |
SR-0740-CS |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product associated with this complaint has not been returned to the manufacturer yet.There was no patient harm reported, however if this malfunction were to reoccur, it potentially would contribute to a serious injury and therefore this complaint will be reported out of abundance of caution.A follow-up mdr will be submitted once additional information is received.Silk road medical will continue to monitor for occurrences of similar events.
|
| |
|
Event or Problem Description
|
|
It was reported that during a tcar procedure, the tip of the stent delivery system (sds) was stuck in the lesion within the stent.Multiple manipulations and interventions were used to free the tip of the sds and the physician converted to a carotid endarterectomy (cea) procedure to resolve the issue.The procedure was completed successfully with no health consequences or impact to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device investigation results did not identify any out-of-specification characteristics or product problems.Procedural and patient factors may have contributed to the withdrawal difficulties.
|
| |
|
Event or Problem Description
|
|
This supplemental mdr is being submitted to update the b1 (adverse event or product problem) and h10-device investigation results for the initial report submitted on 02/07/2023.
|
| |
|