| Model Number |
386100 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the screen is black/pixelated and will not turn on.When it is powered off the screen flashes white and goes completely dark.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) was reviewed and was no failures associated with the reported condition observed.The manufacturing processes were reviewed and there were no abnormalities observed.The unit passed the test and visual inspection prior to releasing for shipment.The device was returned for investigation.An analysis was performed on the returned unit.The unit passed the physical appearances as there were no cosmetic issues found.Furthermore, there were no abnormalities found on the returned unit as the unit was able to power on the screen and start up to ¿language¿ mode.The console was able to login and passed the functional and inspection test.The root cause was not able to determined no failure was observed during the analysis.No corrective actions are needed at this time.This complaint will be used for tracking and trending purposes.
|
| |
|