| Model Number |
FT3 G3 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys ft3 iii results for 1 patient sample on a cobas e 801 module compared to an acuraseed wako analyzer and an abbott architect analyzer.This medwatch will cover ft3 iii.Refer to medwatch with patient identifier (b)(6) for information on the ft3 iii ver.2 results.Refer to the attachment to the medwatch for all patient data.No questionable results were reported outside of the laboratory.The investigational e801 analyzer serial number is (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The patient sample was provided for investigation.Further investigation of the patient sample confirmed an interfering factor against the idiotype of the monoclonal antibody of the ft3 assay.This interference caused the high ft3 results.This interference is covered in product labeling.Per product labeling, "in rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur.These effects are minimized by suitable test design.For diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings." the investigation did not identify a product problem.
|
| |
|