• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR SILTEX ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR SILTEX ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Model Number 3542680
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 12/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a 44-year-old caucasian female who underwent a breast augmentation revision with a mentor siltex round moderate profile, 475cc saline breast implant experienced right-sided deflation post procedure.The issue was diagnosed through office visitation.As a result, patient underwent removal on (b)(6) 2023.
 
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Device evaluation completed on february 28 , 2023: the product was returned to mentor for evaluation.Mentor then conducted a visual inspection, leak testing, and microscopic examination of the returned device.Visual analysis of the returned sample revealed that no damage or anomalies were observed on the sal siltex rnd diap pkg 475cc breast implant.Leak testing was performed, in accordance with mentor procedures, and a tear was noted on the posterior view measuring approximately 0.2 cm.A microscopic examination was performed and the cause of the deflation could not be identified.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: causes of deflation of saline implants include but are not limited to the following events: crease fold failure, intraoperative or postoperative trauma, excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma such as that which can occur during vigorous exercise, athletics, etc.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Additional information received on (b)(6) 2023 indicated that the patient underwent bilateral replacements as follow: l/ cat:3501690, sn: (b)(6), r/ cat: 3501690, sn: (b)(6).Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SILTEX ROUND MODERATE PROFILE
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key16315449
Report Number1645337-2023-01363
Device Sequence Number5898353
Product Code FWM
UDI-Device Identifier00081317001836
UDI-Public00081317001836
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2016
Device Model Number3542680
Device Catalogue Number3542680
Device Lot Number6657063
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/18/2023
Supplement Date Received by Manufacturer02/22/2023
02/28/2023
05/05/2023
Initial Report FDA Received Date02/07/2023
Supplement Report FDA Received Date02/27/2023
03/03/2023
05/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/28/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age44 YR
Patient SexFemale
Patient EthnicityHispanic
Patient RaceWhite
-
-