|
See d4.Lot number and expiration date, d6a.And d10.The reason for this revision surgery the agent reported "(humeral component as well condyle kit was removed and replaced)." the previous surgery and the surgery detailed in this event occurred 5.4 years apart.The actual length of in-vivo for the item(s) listed is unknown as the original surgery date was not provided or could be established.Patient previously had revision surgery due to dislocation.This evaluation is limited in scope as the item(s) associated with this investigation was not returned to djo surgical - austin for examination.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.The findings did not lead to a firm conclusion since the needed records were not made available at the time of this investigation.If more information is received later, the complaint will be updated.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.A review of the device history record(s) show that the reported component(s) used in the previous surgery, when released for use, met design and manufacturing requirements.There were no nonconforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device(s) showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to the components being replaced.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.
|