Model Number 6260-9-044
Medical Device Problem Codes Degraded (1153); Material Erosion (1214); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Metal Related Pathology (4530)
Date of Event 01/20/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2013 and was revised on (b)(6) 2021.It is further alleged that he suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in his blood.
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Additional Manufacturer Narrative
Reported event: an event regarding disassociation and abnormal ion level/metallosis involving a metal head was reported.The disassociation event was confirmed via clinician review of the provided medical records.Abnormal ion level/metallosis was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical records by a clinical consultant indicated: "mechanical failure of a tja is confirmed is confirmed.High cobalt and chrome serum levels cannot be confirmed as no documentation was provided to support this claim.The root cause of the tha mechanical failure cannot be identified from this limited documentation." product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to disassociation of the head from the stem and metallosis/elevated metal ions.A review of the provided medical records by a clinical consultant indicated the following: "mechanical failure of a tja is confirmed is confirmed.High cobalt and chrome serum levels cannot be confirmed as no documentation was provided to support this claim.The root cause of the tha mechanical failure cannot be identified from this limited documentation." no further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2013 and was revised on (b)(6) 2021.It is further alleged that he suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in his blood.Update on 13-april-2023: per revision operative report, the femoral head had disassociated from the stem due to "bird beak" type wear of the stem trunnion.Thick black liquid consistent with metallosis was observed.The stem, head and liner were revised.There were no reported allegations against the liner stated in the op report.The cup was found to be well-fixed so it was not revised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40 COCR LFIT HEAD 44MM/-4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16315830
Report Number0002249697-2023-00138
Device Sequence Number12719799
Product Code JDI
UDI-Device Identifier07613327032710
UDI-Public07613327032710
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K061434
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2016
Device Model Number6260-9-044
Device Catalogue Number6260-9-044
Device Lot NumberMKLE9H
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/18/2023
Supplement Date Received by Manufacturer04/13/2023
Initial Report FDA Received Date02/07/2023
Supplement Report FDA Received Date05/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/25/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age62 YR
Patient SexMale
-
-