• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH PRIMARY TRITANIUM HEMI SOLIDBACK CUP 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH PRIMARY TRITANIUM HEMI SOLIDBACK CUP 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 500-03-52D
Medical Device Problem Codes Loss of Osseointegration (2408); Osseointegration Problem (3003)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 11/04/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient's right hip was revised due to aseptic loosening of the acetabular component.A shell, head, and liner were revised.
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.
 
Additional Manufacturer Narrative
Reported event: an event regarding loosening involving a trident shell was reported.The event was confirmed via clinician review.Method & results: device evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.Clinician review: a review of medical records by a clinical consultant indicated: "this inquiry concerns a patient who developed aseptic loosening of his right acetabular component which required revision to an mdm articulation.I can confirm that the revision was carried out since i was able to review the operation report.The causes of aseptic loosening of an acetabular component several years after surgery are multifactorial and i cannot determine the root cause with certainty.Influencing factors include surgical technique, patient activity level and bmi.With the information provided i would not attribute any causality to the implant itself" device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: a review of medical records by a clinical consultant indicated: "this inquiry concerns a patient who developed aseptic loosening of his right acetabular component which required revision to an mdm articulation.I can confirm that the revision was carried out since i was able to review the operation report.The causes of aseptic loosening of an acetabular component several years after surgery are multifactorial and i cannot determine the root cause with certainty.Influencing factors include surgical technique, patient activity level and bmi.With the information provided i would not attribute any causality to the implant itself".No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported that the patient's right hip was revised due to aseptic loosening of the acetabular component.A shell, head, and liner were revised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMARY TRITANIUM HEMI SOLIDBACK CUP 52MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood WD6 3-SJ
UK   WD6 3SJ
2082386500
MDR Report Key16315859
Report Number0002249697-2023-00141
Device Sequence Number7843694
Product Code LPH
UDI-Device Identifier07613327040043
UDI-Public07613327040043
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K081171
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2017
Device Model Number500-03-52D
Device Catalogue Number500-03-52D
Device Lot NumberMLL2DN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/19/2023
Supplement Date Received by Manufacturer03/15/2023
Initial Report FDA Received Date02/07/2023
Supplement Report FDA Received Date04/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/26/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age75 YR
Patient SexMale
Patient Weight70 KG
Patient EthnicityNon Hispanic
-
-