| Model Number |
8975 |
| Medical Device Problem Codes |
Computer System Security Problem (2899); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A follow up report will be submitted with investigation results should the device be repaired or the device/logs be received for evaluation.Per 803.52(f)(11)(iii) the information provided represents all of the known information at this time.The complainant or reporter was unable or unwilling to provide any further patient, product, or procedural details to the manufacturer.Device was not returned to manufacturing facility.
|
| |
|
Event or Problem Description
|
|
It was reported that product security issues were found during a penetration test on alaris systems manager performed by the customer.The test was performed prior to go-live and there was no patient involvement.
|
| |
|
Event or Problem Description
|
|
It was reported that product security issues were found during a penetration test on alaris systems manager performed by the customer.The test was performed prior to go-live and there was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Omit : c20 - no findings available, d15 - cause not established.
|
| |
|