MEDTRONIC NAVIGATION, INC 2.25MM PEDICLE PROBE; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9733457
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2023
Event Type  malfunction  
Manufacturer Narrative
The probe was returned to the manufacturer for analysis.The tip of the returned probe is visibly bent.Otherwise, with markers attached and fully seated, the probe displays good geometry and divot error readings with normal tracking and verification.Codes b01, c07, and d02 are applicable to this analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system being used during a spinal procedure.It was reported that the instrument had a bent tip.There was no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.25MM PEDICLE PROBE
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key16323661
MDR Text Key309073330
Report Number1723170-2023-00182
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9733457
Device Catalogue Number9733457
Device Lot Number210721
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/21/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-