ARTIVION, INC. ¿ AUSTIN ONX AORTIC CONFORM EXT 27/29; HEART-VALVE, MECHANICAL Back to Search Results
Model Number ONXACE-27/29
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Endocarditis (1834); Insufficient Information (4580)
Date of Event 11/02/2022
Type of Reportable Event Serious Injury
Event or Problem Description
An implant registration card (onxace-27/29 sn (b)(4)) was returned from the user facility for a patient with an existing on-x heart valve (onxace-27/29 sn (b)(4)).Both ircs notate the valves were implanted in the aortic position suggesting the onxace-27/29 sn (b)(4) was explanted.This investigation is relegated to onxace-27/29 sn (b)(4).
 
Additional Manufacturer Narrative
This investigation is currently ongoing.Any additional information will be provided in the follow-up report.This report is being submitted as required by federal regulations and does not constitute an admission that the device caused or contributed to the reported event.Furthermore, this report reflects the event as alleged by the complainant and does not imply that the information reported to artivion ¿ formerly cryolife/jotec is accurate or has been confirmed by artivion ¿ formerly cryolife/jotec.
 
Additional Manufacturer Narrative
A2 - date of birth (b)(6) 1986.G1 - manufacturing site facility name - on-x life technologies, inc.G3/g4 - pma/510(k)number - p000037.H10 - under "additional information narrative" the following statement has been updated to include the reference to the manufacturer."no medical records or operative notes were provided to the manufacturer and the explanted valve was not returned to the [manufacturer] for examination.".
 
Additional Manufacturer Narrative
According to initial reports, an implant registration card (onxace-27/29 sn (b)(6) was returned from the user facility for a patient with an existing on-x heart valve (onxace-27/29 (b)(6) both ircs notate the valves were implanted in the aortic position suggesting the onxace-27/29 (b)(6) was explanted.This investigation is relegated to onxace-27/29 (b)(6) additional information received.According to the surgeon, "the valve was explanted for infection- no concern for valve defect itself.Patient is alive.Valve went to path.Path was consistent with endocarditis with an abscess.Grew staph aureus." the product will not be returned to the manufacturer for evaluation.The manufacturing records for the onxace-27/29 (b)(6) were reviewed and it was confirmed that all records were controlled, available for review, and met all specifications per the device master record.All lots passed functional testing and met release specifications.A review of the available information was performed.An onxace 27/29 (b)(6) was implanted (b)(6) 2022 in a 35-year-old male with an unknown medical history.According to mandatory tracking information received by the manufacturer, this valve was explanted (b)(6) 2022 (28 days post-implant) and replaced with onxace 27/29 (b)(6) additional information from the surgeon indicated that the valve had been removed due to prosthetic valve endocarditis.¿the valve was explanted for infection -no concern for valve deficit itself¿ and ¿path was consistent with endocarditis with an abscess.Grew staph aureus¿.No medical records or operative notes were provided to the manufacturer and the explanted valve was not returned to the for examination.A review of the original manufacturing records shows no abnormalities.With the information available, the source of the endocarditis is unknown.However, because the on-x valve manufacturing process includes validated terminal sterilization prior to distribution¿verified by a review of manufacturing records unique to this valve¿the valve is unlikely to be the source of infection.The instructions for use (ifu) for the on-x valve states that reoperation, including explantation, may result from a complication, in this case, endocarditis, a known potential event acknowledged in the ifu.Though rare, historically, endocarditis occurs at a rate of 0.3 %/patient-year for mechanical aortic heart valves [iso 5840-2:2021].Endocarditis is the root cause for the explantation of the aortic on-x valve.Because all on-x valves undergo validated terminal sterilization prior to distribution, the valve is unlikely to be the source of the infection.No further action is required.Endocarditis is the root cause for the explantation of the aortic on-x valve.Because all on-x valves undergo validated terminal sterilization prior to distribution, the valve is unlikely to be the source of the infection.The manufacturing records were reviewed, and it was confirmed that all records were controlled, available for review, and met all specifications per the device master record.All lots passed functional testing and met release specifications.This event does not identify additional hazards or modify the probability and severity of existing hazards.There is no indication that an error or deficiency occurred at artivion ¿ formerly cryolife/jotec and the ifu adequately communicates risk.This complaint was reviewed for a capa evaluation and a capa is not warranted at this time.Artivion will continue to monitor similar complaints to determine if additional actions are warranted; however, at this time no further actions are necessary.This report is being submitted as required by federal regulations and does not constitute an admission that the device caused or contributed to the reported event.Furthermore, this report reflects the event as alleged by the complainant and does not imply that the information reported to artivion ¿ formerly cryolife/jotec is accurate or has been confirmed by artivion ¿ formerly cryolife/jotec.
 
Additional Manufacturer Narrative
Under "additional information narrative" the following statement has been updated to include the reference to the manufacturer.No medical records or operative notes were provided to the manufacturer and the explanted valve was not returned to the manufacturer for examination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONX AORTIC CONFORM EXT 27/29
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ARTIVION, INC. ¿ AUSTIN
1300 e. anderson ln., bldg. b
austin TX 78752
Manufacturer (Section G)
ON-X LIFE TECHNOLOGIES, INC.
1300 e. anderson ln., bldg. b
austin TX 78752
Manufacturer Contact
rochelle maney
1655 roberts blvd
kennesaw, GA 30144
7704193355
MDR Report Key16324682
Report Number1649833-2023-00003
Device Sequence Number8813472
Product Code LWQ
UDI-Device Identifier00851788001488
UDI-Public851788001488
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
P000037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberONXACE-27/29
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date01/11/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/11/2023
Supplement Date Received by Manufacturer01/11/2023
01/11/2023
01/11/2023
Initial Report FDA Received Date02/08/2023
Supplement Report FDA Received Date03/21/2023
03/29/2023
03/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
Patient SexMale
-
-