ZIMMER BIOMET, INC. PS TIBIAL ARTICULAR SURFACE PROVISIONAL; INSTRUMENT KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/17/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device fractured.Attempts to obtain additional information have been made; however, no more is available.
 
Additional Manufacturer Narrative
(b)(4).Complaint sample was evaluated.Visual evaluation of the returned device shows signs of repeated use and is fractured on the medial side of the post feature.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.A previous investigation into this issue determined that the likely root cause for the tasp fractures is bending/torsional loading on the device.A notification was sent to the field on august 7th, 2014 to provide additional clarifications relating to the instructions for use of the tasp construct for all persona users on file at the time of the field notice.Ps tasp top components and standard (non-cps) tasp bottom components have been modified to prevent fracture.The fractured tasp component in this complaint was manufactured before the design modification.The root cause is considered to be a previously addressed design issue.The event is being reviewed through the capa process.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PS TIBIAL ARTICULAR SURFACE PROVISIONAL
Common Device Name
INSTRUMENT KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16326584
Report Number0001822565-2023-00285
Device Sequence Number20831217
Product Code JWH
UDI-Device Identifier00889024238800
UDI-Public(01)00889024238800(10)62836516
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42517400510
Device Lot Number62836516
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/17/2023
Initial Report FDA Received Date02/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/17/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2578-2014
Patient Sequence Number1
Patient SexUnknown
-
-