| Model Number |
100014514 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During device preparation, with the patient on the table, the ensite amplifier did not pass the self-test and the procedure was cancelled.Even if it succeeded and the lamp was green, it blinked orange after a few seconds.Troubleshooting involved checking all connections several times, restarting the amplifier and the dws with and without cables several times and checking the air filter, which was clean and the procedure was cancelled.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported self test issue and subsequent cancellation remains unknown.
|
| |
|