| Model Number |
45043 |
| Medical Device Problem Codes |
Display or Visual Feedback Problem (1184); Defective Device (2588); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was cancelled.An angiojet ultra 5000a console was selected for use.During preparation, the unit was leaking and the drawer was suspected to be defective.The procedure was cancelled due to this event.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was cancelled.An angiojet ultra 5000a console was selected for use.During preparation, the unit was leaking and the drawer was suspected to be defective.The procedure was cancelled due to this event.No patient complications were reported.It was further reported that error 69 and 43 occurred.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was cancelled.An angiojet ultra 5000a console was selected for use.During preparation, the unit was leaking and the drawer was suspected to be defective.The procedure was cancelled due to this event.No patient complications were reported.It was further reported that error 69 and 43 occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: angiojet drawer was received.The drawer was decontaminated.A visual inspection was done on the drawer, and it failed because it is missing 2 screws.Then a functional test was done on the drawer.It passed the functional test.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: angiojet console (b)(6) was also received and decontaminated.The console failed visual inspection because one of the bag holder assemblies was broken.Then a functional test was done on the console.It failed the functional test by booting up to error code 56.This resulted in product failure, confirming the complaint.Device evaluated by mfr: angiojet drawer was received.The drawer was decontaminated.A visual inspection was done on the drawer, and it failed because it is missing 2 screws.Then a functional test was done on the drawer.It passed the functional test.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was cancelled.An angiojet ultra 5000a console was selected for use.During preparation, the unit was leaking and the drawer was suspected to be defective.The procedure was cancelled due to this event.No patient complications were reported.
|
| |
|