• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL VLOCK ML CLIP 6/CART 14/BOX; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL VLOCK ML CLIP 6/CART 14/BOX; CLIP, IMPLANTABLE Back to Search Results
Model Number IPN055290
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn#(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event or Problem Description
Reported event: clips broke when they were trying to load them.
 
Event or Problem Description
Reported event: clips broke when they were trying to load them.
 
Additional Manufacturer Narrative
(b)(4).On inspection zero defects detected during in-process and final inspections performed by qa.100% inspection of finished goods is performed by in-line inspectors.Nonconforming packages or product is required to be removed from the lot.No component issues were detected during sub-assembly in-process.100% inspection of each cartridge assembly is performed as part of the production process.Nonconforming cartridge assemblies are required to be removed from the lot.Clip inspection of (b)(4) visually showed zero visual defects / zero dimensional nonconformances detected.The ifu informs the user, "to load the applier, grasp the applier and carefully insert the jaws of the applier into the cartridge slot, making sure the jaws are perpendicular to the base of the cartridge.Gently press the applier over the clip until there is an audible click.Do not force the applier into the cartridge or onto the clip.The applier should enter and withdraw from the cartridge easily." corrective action is not required at this time as the probable cause could not be determined based upon the information provided and without a sample.Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VLOCK ML CLIP 6/CART 14/BOX
Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
effie jefferson
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key16328869
Report Number3011137372-2023-00013
Device Sequence Number18304025
Product Code FZP
UDI-Device Identifier24026704666823
UDI-Public24026704666823
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K152082
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2025
Device Model NumberIPN055290
Device Catalogue Number51114V
Device Lot Number228029
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2023
Supplement Date Received by Manufacturer03/01/2023
Initial Report FDA Received Date02/08/2023
Supplement Report FDA Received Date03/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED.; NONE REPORTED.
Patient SexUnknown
-
-