ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458QL/86
Device Problems Capturing Problem (2891); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Event Description
Related manufacturer reference number: 2017865-2023-05439.It was reported that a patient presented in-clinic.Examination of the right ventricular and left ventricular lead revealed dislodgement.Both leads were explanted and replaced on (b)(6) 2023.No adverse patient consequences were reported.
 
Manufacturer Narrative
Further information was requested, but not received.
 
Manufacturer Narrative
The reported events were lead dislodgement and inadequate capture.As received, a complete lead was returned in one piece.The reported event of inadequate capture was not confirmed.Electrical testing did not find any indication of internal shorts.The s-curve height was measured within specification.Visual inspection of the lead did not find any anomalies except for procedural damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16329775
MDR Text Key309121037
Report Number2017865-2023-05440
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510219
UDI-Public05414734510219
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2024
Device Model Number1458QL/86
Device Catalogue Number1458QL/86
Device Lot NumberA000108568
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/20/2023
Initial Date FDA Received02/08/2023
Supplement Dates Manufacturer Received02/19/2023
Supplement Dates FDA Received02/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/14/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GALLANT HF; TENDRILLEAD
Patient Outcome(s) Required Intervention;
Patient Age55 YR
Patient SexMale
-
-