• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO DUET®; DEVICE, MONITORING, INTRACRANIAL PRESSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO DUET®; DEVICE, MONITORING, INTRACRANIAL PRESSURE Back to Search Results
Model Number 46914
Medical Device Problem Codes Break (1069); Disconnection (1171); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event occurred in canada.No products have been returned to medtronic for analysis.Are applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding an external drainage and monitoring system (edms) used during an external ventricular drainage in icu.It was reported that the connection was always either breaking or coming loose at the stopcock or at the leurlock connected to the evd catheter.The tubes were either breaking or becoming disconnected "at random," there was no tension on the lines nor was the patient pulling at their lines.It was noted that this was the second occurrence.There was no reported impact to the patient.There was no reported delay to the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUET®
Common Device Name
DEVICE, MONITORING, INTRACRANIAL PRESSURE
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key16330127
Report Number9612164-2023-00567
Device Sequence Number6419530
Product Code GWM
UDI-Device Identifier00613994445377
UDI-Public00613994445377
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K983799
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Model Number46914
Device Catalogue Number46914
Device Lot Number224990659
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2023
Initial Report FDA Received Date02/08/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-