• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; HDVIDEO COLONOSCOPE 2.8C 9.8T 1700L FWJ

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; HDVIDEO COLONOSCOPE 2.8C 9.8T 1700L FWJ Back to Search Results
Model Number EC-2990LI
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2023
Type of Reportable Event Malfunction
Event or Problem Description
The time of event is unknown.There was no report of patient harm.Air/water nozzle peeled off (black glue).
 
Additional Manufacturer Narrative
We checked the returned unit and confirmed that the nozzle gluing missing.Based on the result, we concluded that it was caused due to the physical damage applied on the nozzle gluing.In addition, we confirmed that the insertion flexible tube (ift) coating damage, the light guide fiber bundle (lcb) broken, the insertion flexible tube (ift) cut, the light guide cable bump, the suction channel kink, and the angle wire grinding; however, they are not the main cause, and/or irrelevant to the alleged complaint.In terms of gluing missing, the possibility of dropping into human body could not be denied.Moreover, based on the technical report""hr-rpt-0585(nozzle)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
HDVIDEO COLONOSCOPE 2.8C 9.8T 1700L FWJ
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
MDR Report Key16330872
Report Number9610877-2023-51255
Device Sequence Number6419551
Product Code FDF
UDI-Device Identifier04961333163335
UDI-Public04961333163335
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K131855
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEC-2990LI
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2023
Initial Report FDA Received Date02/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-