MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE PHACO TIP; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE PHACO TIP; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065740837
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported tip was found to be bent when opened.The surgery was performed and completed after replacing the product with another one.There was no report of the patient harm.
 
Additional Manufacturer Narrative
One opened phaco tip, bent in a bag, was received for the report.The phaco tip was visually inspected and deemed nonconforming.The cannula is bent at the cone to cannula transition area.There was wear on the threads and back of flange that was consistent with threading on a handpiece.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The complaint evaluation confirms the phaco tip cannula is bent.How and when the phaco tip cannula became bent cannot be determined from this evaluation.The most likely cause of a bent phaco tip cannula is from improper handling of the product.No specific action with regard to this complaint was taken because the root cause for the complaint issue cannot be determined from this evaluation.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Any nonconformance, such as the phaco tip cannula bent exhibited on the returned opened sample, is removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PHACO TIP
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16331748
Report Number2523835-2023-00052
Device Sequence Number6789524
Product Code HQC
UDI-Device Identifier00380657408375
UDI-Public00380657408375
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K161794
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065740837
Device Lot Number14DYE5
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/03/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/12/2023
Supplement Date Received by Manufacturer04/06/2023
Initial Report FDA Received Date02/09/2023
Supplement Report FDA Received Date05/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/24/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-