| Model Number |
CB811 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that prior to use of a fusion oxygenator, the customer reported a temp port leak and prior to the use of a custom tubing pack, the customer reported a kink in the green gas tubing.The device was replaced.There was no patient involvement, so no adverse effect occurred.Medtronic received additional information that there was no damage to the packaging.The kink was found in the sterile packaging.There was no visible air in system/tubing.The kink in the custom tubing pack was discovered straight out of the packaging.The leak in temp port was discovered during priming.A perfusionist inserted the temp probe and then the perfusionist who was pumping the case did not like something about the setup.They removed the temp probe and noticed a leak after removal.There was no additional manipulation of the port.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic received additional information that the leak was noticed after the temperature probe was removed and inserted again.Device evaluation: visual inspection showed that the temperature monitoring adapter (tma) was missing from the device.When the device was being rinsed out there was a leak from the tma port.The reason for return was confirmed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the complaint is confirmed for the fusion oxygenator's loose and leaking temperature monitoring adapter (tma).The issue was verified via analysis of the returned device.Review of this unit¿s device history record found no abnormalities or non-conforming material reports (ncmrs) initiated during manufacturing that would cause or contribute to the reported event.This unit passed all required testing and inspections during manufacturing.Customer communication of this issue has been initiated via fa1302, which provides device use recommendations for affected product.Root cause investigation concluded the presence of ebs (ethylene bis-stearamide) wax on the new tma component affected the material properties that caused the torque specification to not be met.There were no patient/clinical safety issues reported.Trends for issues with this product are reviewed at quarterly quality meetings.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|