| Catalog Number |
GEN11 |
| Medical Device Problem Codes |
Failure to Read Input Signal (1581); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unknown procedure the unit emits a burning smell and fails the electrical safety test.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date of event: event year only reported: 2023.Batch #: unknown.The lot/batch was not provided; therefore, the manufacturing records evaluation could not be performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 2/15/2023.Investigation summary: per service manual operational and diagnostic analysis confirmed the burning smell and the electrical safety failure.The epac top, epac middle, epac bottom and cable assembly, enclosure / heat sink fan to main board were replaced as identified in the investigation to address the issues.The repair and testing of the unit was completed per the service manual, bringing the unit back to full functionality.The unit passed all functional tests and is fully operational.The repair and testing of the unit was completed.The device history records were reviewed and certed by external manufacturing that the manufacturing criteria were met prior to the release of the equipment.The certificate records are accessible through external manufacturing.
|
| |
|
Search Alerts/Recalls
|