Model Number 1488210105
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/23/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was blurry image.
 
Additional Manufacturer Narrative
Alleged failure: blurry grainy image.Confirmed failure: chd-a-lcj loose cable jacket probable root cause: ¿ cables ¿ connectors ¿ digital board ¿ power supply ¿ filter / fuse ¿ ac inlet board ¿ main board ¿ transition board ¿ fiber board ¿ intermittence between transition board and ch connector ¿ software ¿scope (see rsk10356 output ""visualization/image"") ¿light source (see rsk10975 output ""white light"") ¿ camera head (sensor, sensor cable) ¿ coupler (dust on optics, scratched optics, broken / shifted lenses, focusing mechanism, endoclamp, zoom ring) ¿ 1488 & 1588 extension cable ¿ intermittence while using rf devices (ex: valleylab) ¿ problem toggling light source or from camera head ¿ connected monitors ¿ leaking ¿ use error ¿ poor grounding (e.G.""finger"" grounding tab connecting front and rear enclosure.Camera head connection from cable to transition board.) ¿electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge ¿ manufacturing/ service nonconformity" the product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that there was blurry image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 1488 HD 3-CHIP CAMERA HEAD, C-MOUNT
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key16342759
Report Number0002936485-2023-00111
Device Sequence Number6882153
Product Code GCJ
UDI-Device Identifier07613327061055
UDI-Public07613327061055
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K132785
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1488210105
Device Catalogue Number1488210105
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/23/2023
Supplement Date Received by Manufacturer01/23/2023
Initial Report FDA Received Date02/09/2023
Supplement Report FDA Received Date08/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-