• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. HIP SYSTEMS DJO SURGICAL; HEAD, FEMORAL, COCR 28/-3.5MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. HIP SYSTEMS DJO SURGICAL; HEAD, FEMORAL, COCR 28/-3.5MM Back to Search Results
Model Number 497-28-035
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 01/24/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - infection.
 
Additional Manufacturer Narrative
The agent reported "(patient had m/m and had an infection)".The previous surgery and the surgery detailed in this event occurred 18 years apart.This evaluation is limited in scope as limited information was provided to djo surgical - austin for examination.If information regarding cultures identified in the infection, the severity of the infection or any other relevant information is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.The root cause of this complaint was a revision surgery due to an infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.
 
Additional Manufacturer Narrative
Corrected data.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIP SYSTEMS DJO SURGICAL
Common Device Name
HEAD, FEMORAL, COCR 28/-3.5MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, TX 78758-5445
MDR Report Key16343415
Report Number1644408-2023-00172
Device Sequence Number17721724
Product Code LPH
UDI-Device Identifier00888912087278
UDI-Public00888912087278
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K932481
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2010
Device Model Number497-28-035
Device Catalogue Number497-28-035
Device Lot Number210682
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2023
Supplement Date Received by Manufacturer02/13/2023
Initial Report FDA Received Date02/09/2023
Supplement Report FDA Received Date02/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/24/2004
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
432-28-207 , LOT 213652.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
-
-