| Model Number |
DWD182 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a glenosphere defect.Another implant was used to treat the patient.No impact on the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed (jammed safety screw in the glenoid sphere after opening the package) since the device was returned and was found to be in condition stated in the event description.The device inspection revealed the following: the visual inspection confirmed that the safety screw was jammed in the glenoid sphere.After loosening the safety screw, the functionality of the sphere was checked, and it could correctly impact and tighten into the baseplate-gauge without issue to report.The issue described in the reported event was unable to be reproduced.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a glenosphere defect.Another implant was used to treat the patient.No impact on the patient.
|
| |
|