• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number DFW225
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Disturbances (2140)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the dfw225 intraocular lens (iol) was implanted into the patient¿s ocular dexter (right eye).Patient reports complaints of inability to drive at night within days of first surgery.There was no patient injury.Iol explant is scheduled for (b)(6) 2023.No further information is available.
 
Additional Manufacturer Narrative
Additional information: patient weight and ethnicity and race: unknown/asked but unavailable.Date of event: unknown/asked but unavailable.Best estimate date provided as days within first surgery.Date explanted: not applicable, as the iol remains implanted.The device was not returned for evaluation as it remains implanted; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts were made to contact the customer requesting additional information regarding complaint, however, to date no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Correction: device was returned thus indicating the iol was explanted.The following fields were updated accordingly: section d-6b date explanted: unknown, information not provided.Previously reported in the initial report as n/a as device remains implanted.Section h-6 health effect - impact code: updated to 4629 - device revision or replacement, previously reported in the initial report as 2199 - no health consequences or impact.Additional information: section d-9: device available for evaluation: yes.Section d-9: device date returned to manufacturer: 14-aug-2023.Section h-3: device evaluated by manufacturer: yes.Device evaluation: one lens was received inside of a specimen cup marked with two sn's.Sn:(b)(6) and sn: (b)(6).Since it cannot be determined which investigation the lens belongs to, it will be evaluated in both investigations.Visual inspection under magnification revealed that the complaint lens was received cut in half.The lens was cleaned and, no issues that could cause or contribute to the complaint issue.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue was not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Conclusion: based on the complaint investigation results, the product was released within specifications.The complaint issue reported could not be confirmed and no product deficiency or product malfunction could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key16344478
Report Number3012236936-2023-00213
Device Sequence Number6319013
Product Code MFK
UDI-Device Identifier05050474751781
UDI-Public(01)05050474751781(17)241216
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/16/2024
Device Model NumberDFW225
Device Catalogue NumberDFW225U165
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2023
Supplement Date Received by Manufacturer09/25/2023
Initial Report FDA Received Date02/09/2023
Supplement Report FDA Received Date10/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-