During a tavr, (transaortic valve replacement) was being deployed into the aortic valve, at optimal inflation, the balloon rupture occurred during deployment of the valve.The valve was deployed successfully.Since the balloon was ruptured, there was difficulty removing the delivery system through the sheath.The sheath and the edwards delivery system had to be removed simultaneously.There were minor complications at the insertion site which were resolved with percutaneous transluminal angiogram (pta) of the right common femoral artery.The patient was subsequently transferred to the pacu (post anesthesia care unit) in stable condition.Manufacturer response for edward lifescience sapien 3 delivery system, sapien 3 delivery system (per site reporter).Manufacturer rep is always present during this procedure.Was present when balloon ruptured.
|