Model Number A219
Device Problems No Audible Alarm (1019); Interrogation Problem (4017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2023
Event Type  malfunction  
Event Description
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) was not responsive to interrogations after using numerous programmers, and to magnet application.Beeping tones were not emitted when expected.It was suspected that this issue was caused as the patient had a magnetic resonance imaging (mri) and the mri mode of the s-icd was not turned on.Reportedly, the device never regained any functionality.The patient has other more critical care needs that are being addressed, and a replacement procedure while occur until further notice.This s-icd remains in service and no adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM MRI S-ICD
Type of Device
SUBCUTANEOUS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (S-ICD)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16353402
MDR Text Key309293894
Report Number2124215-2023-05956
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526584404
UDI-Public00802526584404
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110042/S058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/22/2022
Device Model NumberA219
Device Catalogue NumberA219
Device Lot Number129052
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/20/2023
Initial Date FDA Received02/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/03/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age56 YR
Patient SexFemale
-
-