| Model Number |
20161420-1 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
Burn(s) (1757)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.The event is filed under internal karl storz complaint id: case no.(b)(4).
|
| |
|
Event or Problem Description
|
|
It has been reported the staff has received burns from the light lead caused by this device.
|
| |
|
Additional Manufacturer Narrative
|
|
During inspection at our repair facility a charred fiber taper was detected.This is assumed to be caused either by a dirty or unsuitable fiber optic cable.Due to the contamination or unsuitable light guide, the light cannot be guided sufficiently to the outside and the fiber taper heats up so much that it gets charred.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is to correct section d9 to reflect the correct device return date and section h6 conclusion code for the evaluation.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|