• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL CT BRD VIO 27IN 0 S/A FSL; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL CT BRD VIO 27IN 0 S/A FSL; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number J587H
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Component code: g07002 - device not returned.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: did the two devices returned are just samples or did this devices present the issue reported? please confirm.The two devices returned are each a sample from the two boxes of suture that have the needles in question.Rdbbjjto is the lot number for one box of suture and sgbdqdto is the lot number for a second box of suture.Both boxes of suture have the same type of needles.The issue is that the needle starts off as a cutting needle at the tip, but then becomes a taper needle further back.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.This complaint was initially reported for product code j587h and lots sgbdqdt0 and rdbbjjt0.However our failure analysis lab also received a wrong product with reference to this complaint, we received the following product: product code: j195.Product lot #: mg7158.1.Could you please clarify if product code j195 and lot mg7158 belongs to this complaint? 2.If not, could you please clarify if the wrong received product belongs to a different complaint? if so, can you please provide the complaint number.3.If the device was not reported before, please provide more details of the alleged quality issue.4.If the product doesn¿t belong to any complaint, please let us know so we can discard it or advise if the product is required to be returned.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Event related to mw # 2210968-2023-00974.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that an animal underwent an unknown procedure on an unknown date and suture was used.The account reported that the suture they ordered was labeled as having a ¿cutting needle¿ but when the package was opened for a procedure it was noticed that the needle attached to the suture was a ¿tapered needle¿.The same device was used to complete the procedure.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Investigation summary:the product was returned for evaluation.The returned product determined that it was received one unopened sample that pertains to product code j587h.In order to evaluate the condition of the returned sample, the packet was opened.The swage and attachment area were noted to be as expected.The tip and body needle were inspected, and no anomalies or issues related to product mix were observed during evaluation.In addition, the needle was compared against needle specification and met with the specification.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 3/30/2023.Additional information was requested, the following was obtained: as mentioned on the initial declaration sheet, product j 195 lot mg7158 ( item #3/needle only) was enclosed as an example of what type of needle we expect to find on reverse cutting suture.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2023-00974 & 2210968-2023-00975.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL CT BRD VIO 27IN 0 S/A FSL
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-GMBH
p.o. box 1409
norderstedt D2284 1
GM   D22841
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key16356269
Report Number2210968-2023-00975
Device Sequence Number12329733
Product Code GAM
UDI-Device Identifier10705031041486
UDI-Public10705031041486
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K022269
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberJ587H
Device Catalogue NumberJ587H
Device Lot NumberSGBDQDT0
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2023
Supplement Date Received by Manufacturer02/15/2023
03/09/2023
Initial Report FDA Received Date02/10/2023
Supplement Report FDA Received Date03/06/2023
03/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-