| Model Number |
RA002-4545SLR |
| Medical Device Problem Codes |
Degraded (1153); Structural Problem (2506); Failure to Advance (2524); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An investigation report from an implant retrieval center was received and reported that during examination of the rod, the retrieval center found the rod exerted 32.8lbf which does not meet the specification.Additionally, the radial bearing was seized to the magnet and the rod could not be further disassembled, therefore, the drive pin was unable to be analyzed.It was additionally reported that the o-ring seal was broken and misshapen.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected section h6- device code.
|
| |
|
Event or Problem Description
|
|
No additional information provided.
|
| |
|