MAUDE Adverse Event Report: MEDTRONIC IRELAND VALIANT NAVION; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND VALIANT NAVION; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR Back to Search Results
Model Number VNMF3434C174TU
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cardiac Arrest (1762); Shock (2072); Pericardial Effusion (3271); Liver Failure (4492); Insufficient Information (4580)
Date of Event 02/08/2020
Type of Reportable Event Death
Additional Manufacturer Narrative
Concomitant medical products: other relevant device(s) are: product id: v nmf3434c97tu, serial/lot #: (b)(4), ubd: 04-feb-2021, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Valiant navion stent grafts were implanted in the endovascular treatment of a dissection. it was confirmed the stent graft (b)(4) was implanted proximally.It was reported that during the index procedure, the patient was found to have a intimal defect at the proximal landing site causing a type a dissection and ia endoleak.  a ct was performed approximately 8 months post the index procedure, which showed the type ia endoleak.Another ct  also showed the type ia endoleak and aortic enlargement at the proximal tevar site.  intervention was performed, a non mdt stent graft was placed in the aortic arch.At another intervention date a lsca bypass was performed and a non mdt aortic root graft was placed during a hemi arch replacement.  corelab reviewed ct imaging dated on (b)(6) 2020; the type ia endoleak was noted on the stent graft (b)(4).Ne for fractures on this stent graft due to device overlaps.No fractures were noted otherwise, no stent ring enlargement or stent ring migration.The maximum aortic diameter was 70.8mm.It was reported the patient died on an unknown date.The cause of death is unknown.  no additional clinical sequelae were provided and the patient is expired.
 
Additional Manufacturer Narrative
Additional information received: it was clarified that the type ia endoleak was observed first on (b)(6) 2020 ct.The dissection and intimal defect along with the persistent type ia endoleak were observed again on ct on (b)(6) 2020.The patient was then transferred emergently for pericardial effusion and visualized dissection flag.A type a dissection repair using non mdt devices was performed.A cardiac arrest subsequently occurred in icu assessed as possibly secondary to rv failure.The patient developed liver failure due to shock and did not recover neurologically.The patient expired on (b)(6) 2020.Per the physician the subacute aortic type a dissection was assessed likely secondary to prior tevar as there was an intimal defect at the proximal tevar landing site.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information received: corelab review of ct imaging from approximately 8 months post-index procedure identified a type ia endoleak and aortic enlargement of +9mm.Corelab re-analyzed the ct imaging at approximately 9 months post index identified aneurysm enlargement of +13.5mm.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
Additional information received : corelab review of ct imaging from approximately 8 months post-index procedure did not identify any fractures, stent ring enlargement or stent ring migration.The maximum aortic diameter was 66.3mm.Corelab review of ct imaging from approximately 9 months post index procedure identified a persistent type ia endoleak.Aortic enlargement was noted both images when compared to imaging from 6 days after the index procedure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT NAVION
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16361017
Report Number9612164-2023-00614
Device Sequence Number20718786
Product Code MIH
UDI-Device Identifier00763000100643
UDI-Public00763000100643
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/14/2021
Device Model NumberVNMF3434C174TU
Device Catalogue NumberVNMF3434C174TU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/21/2023
Supplement Date Received by Manufacturer04/20/2023
09/27/2023
09/27/2023
Initial Report FDA Received Date02/13/2023
Supplement Report FDA Received Date05/04/2023
09/29/2023
10/04/2023
Date Device Manufactured01/15/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10."
Outcome Attributed to Adverse Event Death; Required Intervention;
Patient Age72 YR
Patient SexMale