We checked the returned unit and confirmed that the biopsy inlet t-piece clogged.Based on the result, we concluded that it was caused due to the insufficient reprocessing at the facility on the biopsy inlet t-piece.In addition, we confirmed that the insertion flexible tube (ift) perforated, the light guide cable for prong perforated, and the suction cylinder control body clogged; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0588(channel)"" and/or the risk analysis results, it was evaluated to submit mdr.
|